Skip to main content
Fig. 7 | Journal of Intensive Care

Fig. 7

From: Fundamental concepts and the latest evidence for esophageal pressure monitoring

Fig. 7

Method of calculating WOB. Permitted to reprint from [36] with permission form Springer Nature. x-axis: esophageal pressure, y-axis: lung volume. Since the patient's Work of breathing (WOB) includes the work of expanding the thorax in addition to the work of expanding the lungs, WOB is the area bounded by the straight line whose slope is the chest wall compliance and the esophageal pressure (x-axis) and lung capacity (y-axis) plot curves in the inspiratory phase [21]